Table of Contents |
---|
...
Fick's law of diffusion: ......
being edited
Polarographic Electrode
A typical polarographic electrode consists of a silver anode, a gold or platinum cathode and an electrolyte solution (KCl or AgCl). In order to create a sensor, a constant voltage of 0.8 volts is applied to the probe, and a digital meter is installed to read the DO response measured by the sensor.
...
- Since the electrode consumes oxygen, readings are affected by flow across the sensor tip. Thus enough flow rate at the membrane (or sample renewal rate) must be ensured for accurate results.
Galvanic Electrode
Measuring dissolved oxygen with either sensor type
...
- Winkler titration
- Air-saturated water
- Water-saturated air
Cleaning and Maintenance
[need to be paraphrased]
...
Display of the concentration: The instrument requires the appropriate data of the calibration curve and uses them to calculate the concentration in mg/L (ppm is identical in this case), allowing for the temperature dependency of the individual parameters
Display of the percentage of oxygen saturation: The instrument measures the sensor current and calculates the partial pressure of oxygen according to the calibration. The current air pressure is measured for the calculation of the saturation partial pressure. The display corresponds to the quotient, converted into a percentage.
References
Erlich Industrial Development. D. O. theory. http://www.eidusa.com/Theory_DO.htm
...