Introduction
Dissolved Oxygen (DO) is an essential measurement parameter in aerobic bioreactors. The growth of all cells is heavily dependent on DO because it acts as a terminal electron acceptor in aerobic respiration. However, if excessive amount of DO is added to the process, it may limit the growth of the culture and promote undesirable organisms. Consequently, the measurement of DO is critical to effective operation of systems. Today, a variety of sensors are available in the market, each with its own advantages and disadvantages.
Optical Sensors
Electrochemical Sensors
Electrochemical DO electrodes are divided into two separate types: polarographic and galvanic. These electrodes are constructed with an anode and a cathode submerged in an electrolyte solution. An oxygen-permeable membrane is used to confine the cathode. When the cathode is polarized with a constant voltage, dissolved oxygen molecules diffusing through the membrane is reduced at the cathode. Then, an electrical signal produced by the cathode travels to the anode and then to the instrument. The oxygen tension versus the electrode current can be calibrated since the diffusive flux is a function of the partial pressure of oxygen in the flow. Instead of measuring the DO concentration, the electrode senses DO activity (or tension).
Reduction reaction at the cathode: O2 + 4e + 4H+ → 2H2O
As in the case for the polarographic electrodes, a voltage is applied externally while an internal potential is generated as in the galvanic electrodes.
Polarographic Electrode
A typical polarographic electrode consists of a silver anode, a gold or platinum cathode and an electrolyte solution (KCl or AgCl). In order to create a sensor, a constant voltage of 0.8 volts is applied to the probe, and a digital meter is installed to read the DO response measured by the sensor.
(being edited)
Limitations
- Response time is described as the time required for the electrode to reach >90% of the output. Typical response time for polarographic sensors are is 30 sec, which makes them not compatible to be used for dynamic measurements.
- Warm-up time for this type is approximately 10 minutes. Wrong readings will occur if measurements are made when the required amount of time has not been attained.
- Chloride ions in the electrolyte will be eventually consumed resulting in gradual drift in the electrode signal. The electrolyte must be replaced.
- Since the electrode consumes oxygen, readings are affected by flow across the sensor tip. Thus enough flow rate at the membrane (or sample renewal rate) must be ensured for accurate results.
Galvanic Electrode
References
Contributors:
Eila Myllylä, Chi-En Shan, Olga Gerasimenko, Judit Nyári & Nhat Nguyen